- Color & Paint Coating Measurement
- Dimensional Measurement
- Electrical & Electronic Meter
- Electrostatics (ESD) , Ionizer & Resistivity Meter
- Environmental & Weather Aging Testing
- Gas Measurement
- Handling & Cutting Tools
- Hardness Testing
- Laboratory Equipments
- Levelling Angle Gauges
- Magnetic Testing
- Medical Equipment
- Metallography
- Microscope, Magnifier & Visual Inspection
- Paper & Film Packaging Testing
- Pressure & Flow
- Rubber & Glove Testing
- Storage Dry Cabinet
- Temperature, Humidity & Moisture
- Tensile & Tension Torque/Force
- Ultrasonic NDT Testing
- Vise & Holder
- Water Analysis Meter
- Weighing
- Others
FLUKE - T6-600 ELECTRICAL TESTER
Previous | 367 / 503 | Next |
FLUKE - T6-600 ELECTRICAL TESTER
Brand : Fluke
Model : T6-600
Product Description
Working in cramped electrical boxes and trying to find a safe metallic contact points is often challenging. The Fluke T6-600 Electrical Tester with FieldSense™ technology, lets you quickly and safely take measurements by sliding the open fork over a conductor, without the need for test leads. The T6-600 works with wires up to AWG 4/0 (120 mm2) with a 17.8 mm jaw opening.
Technical Specification
Brand : Fluke
Model : T6-600
Product Description
Working in cramped electrical boxes and trying to find a safe metallic contact points is often challenging. The Fluke T6-600 Electrical Tester with FieldSense™ technology, lets you quickly and safely take measurements by sliding the open fork over a conductor, without the need for test leads. The T6-600 works with wires up to AWG 4/0 (120 mm2) with a 17.8 mm jaw opening.
Technical Specification
Function | Requires test leads | Range | Resolution | Accuracy[1] |
---|---|---|---|---|
FieldSense ac voltage true-rms |
No | 600 V | 1 V | ±(3% + 3 counts) 45 Hz - 66 Hz[2] [3] |
FieldSense ac current true-rms |
No | 200.0 A | 0.1 A | ±(3% + 3 counts) 45 Hz - 66 Hz |
FieldSense frequency (Hz) |
No | 1 Hz | ±(1% + 2 counts)[3] | |
Volts ac true-rms | Yes | 600 V | 1 V | ±(1.5% + 2 counts) 45 Hz - 66 Hz |
Volts dc | Yes | 600 V | 1 V | ±(1% + 2 counts) |
Resistance | Yes | 2000 Ω | 1 Ω | ±(1% + 2 counts) |
Safety | 600 V CAT III | |||
Accuracy[1]: ± ( [ % of reading] + [number of least significant digits] ). Accuracy is specified for 1 year after calibration, at 18 °C to 28 °C (64 °F to 82 °F) with relative humidity to 90 %. AC measurements are ac-coupled, RMS responding. Accuracy[2]: Add 3 % without an external ground connection. External ground connection required for user wearing insulated gloves, standing on an insulated ladder, or otherwise insulated from earth ground. Accuracy[3]: FieldSense is specified from 16 V to 100 % of range. |
Calibration | 1-year calibration cycle | |
---|---|---|
Dimensions | 61.1 x 259.3 x 43.8 mm | |
Jaw opening | 17.8 mm | |
Weight | 0.35 kg (0.78 lb) | |
Temperature | Operating | -10 °C to +50 °C (14 °F to 122 °F) |
Storage | -30 °C to +60 °C (-22 °F to +140 °F) | |
Altitude | Operating | 2,000 m (6,562 feet) |
Storage | 10,000 m (32,808 feet) | |
Relative Humidity | 0% to 90%, 5 °C to 30 °C (41 °F to 86 °F) | |
0% to 75%, 30°C to 40 °C (86 °F to 104 °F) | ||
0% to 45%, 40°C to 50 °C (104 °F to 122 °F) | ||
Battery Type and Life | 2 x AA (IEC LR6); 360 hours continuous, typical; 200 hours using FieldSense | |
Temperature coefficient | 0.1 x (specified accuracy) / °C for <18°C or >28 °C | |
(< 64.4 °F or > 82.4 °F) | ||
Safety | IEC/EN 61010-1: Pollution Degree 2 | |
Electromagnetic Environment | IEC/EN 61326-1: Portable |