- Color & Paint Coating Measurement
- Dimensional Measurement
- Electrical & Electronic Meter
- Electrostatics (ESD) , Ionizer & Resistivity Meter
- Environmental & Weather Aging Testing
- Gas Measurement
- Handling & Cutting Tools
- Hardness Testing
- Laboratory Equipments
- Levelling Angle Gauges
- Magnetic Testing
- Medical Equipment
- Metallography
- Microscope, Magnifier & Visual Inspection
- Paper & Film Packaging Testing
- Pressure & Flow
- Rubber & Glove Testing
- Storage Dry Cabinet
- Temperature, Humidity & Moisture
- Tensile & Tension Torque/Force
- Ultrasonic NDT Testing
- Vise & Holder
- Water Analysis Meter
- Weighing
- Others
HIOKI - Professional pH/mV Meter with Built-in Printer HI122-02
Previous | 29 / 71 | Next |
HIOKI - Professional pH/mV Meter with Built-in Printer
Brand: Hioki
Model: HI122-02
The HI122 pH/mV and temperature benchtop meter feature CAL Check™, a clear and informative LCD display, data logging capability, and computer connectivity. The pH Meter with Built-in Printer HI122 has a single input channel that allows for the measurement of pH and ORP. Readings for pH can be manually or automatically compensated for temperature variations with the separate HI7662-T temperature probe from -20.0 to 120.0°C.
This pH Meter with Built-in Printer can be calibrated up to five points with a choice of seven standard buffers or two custom buffers. All GLP data regarding pH and relative mV calibration is stored for the user to review when necessary. The built-in impact printer incorporated into the HI122 allows for on demand printing of a complete set of information based on the measured, set or recorded data while in various modes such as, measurement mode, GLP or Setup mode.
Features:
pH Specifications:
Temperature Specifications:
Additional Specifications:
Brand: Hioki
Model: HI122-02
The HI122 pH/mV and temperature benchtop meter feature CAL Check™, a clear and informative LCD display, data logging capability, and computer connectivity. The pH Meter with Built-in Printer HI122 has a single input channel that allows for the measurement of pH and ORP. Readings for pH can be manually or automatically compensated for temperature variations with the separate HI7662-T temperature probe from -20.0 to 120.0°C.
This pH Meter with Built-in Printer can be calibrated up to five points with a choice of seven standard buffers or two custom buffers. All GLP data regarding pH and relative mV calibration is stored for the user to review when necessary. The built-in impact printer incorporated into the HI122 allows for on demand printing of a complete set of information based on the measured, set or recorded data while in various modes such as, measurement mode, GLP or Setup mode.
Features:
- Automatic Calibration: pH calibration can be performed up to five points with seven standard buffers and two custom buffers.
- Built-in Impact Printer: The built-in impact printer incorporated into the HI122 uses regular paper that does not fade with time. The information related to measurements being taken can be printed while in measurement mode, GLP or Setup mode. This meter also allows users to print detailed information in four languages for specific help screens and instrument set-up.
- HI1131P pH Electrode: The pH Meter with Built-in Printer HI122 is supplied with the HI1131P glass body, double junction, refillable pH electrode with an indicating sensor made of High Temperature (HT) glass. The double junction and HT glass design allows the HI1131P to be used in a wide variety of applications ranging from samples with metals and Tris buffer to samples at elevated temperatures.
- Temperature Compensation: Temperature for pH measurements can be compensated for automatically (ATC) or manually (MTC) from -20.0 to 120.0°C with the use of the supplied HI7662-T temperature probe.
- GLP Data: The calibration data for each channel including date, time, standards used, offset, and slope can be accessed at any time through the menu of this pH Meter with Built-in Printer.
- Data Logging: The log-on-demand feature accepts the recording of 50 samples. Interval logging allows up to 1000 data points to be recorded and allows the user to specify time intervals from 5 seconds to 180 minutes.
- Data Transfer: With a built-in logging function, measurements are stored in non-volatile memory and can be transferred to a PC through the RS232 port.
pH Specifications:
pH Range | -2.000 to 16.000 pH, -2.00 to 16.00 pH |
pH Resolution | 0.001 pH, 0.01 pH |
pH Accuracy (@25ºC/77ºF) | ±0.01 pH, ±0.002 pH |
pH Calibration | automatic, up to five-point calibration with seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) and two custom buffers |
mV Range | ±999.9 mV; ±2000 mV |
mV Resolution | 0.1 mV; 1 mV |
mV Accuracy | ±0.2 mV (±699.9 mV); ±0.5 mV (±999.9 mV); ±1 mV (±2000 mV) |
Relative mV Offset Range | ±2000 mV |
Temperature Specifications:
Temperature Range | -20.0 to 120.0 ºC, -4.0 to 248.0 °F |
Temperature Resolution | 0.1°C (0.1°F) |
Temperature Accuracy | ±0.4°C (±0.7°F) |
Temperature Compensation | automatic or manual from -20.0 to 120.0 °C |
Additional Specifications:
Temperature Probe | HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) |
Logging | 50 samples (25 per channel), log-on-demand |
Logging Interval | 5 seconds to 180 minutes, 1000 samples (500 per channel) |
Connectivity | RS232 serial port, opto-isolated |
Environment | 0 to 50°C (32 to 122°F), RH max 95% non-condensing |
Power Supply | 12 VDC adapter (included) |
Dimensions | 280 x 203 x 84 mm (11.0 x 8.0 x 3.3’’) |
Weight | 1.9 kg (4.2 lbs.) |
Printer | built-in dot matrix printer, with 44 mm plain paper |
Ordering Information | HI122-02 (230V) are supplied with HI1131P pH electrode, HI7662-T temperature probe, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI7082 3.5M KCL electrolyte solution (30 mL), (5) paper rolls, 12 VDC adapter and instructions. |