CEDAR - Digital Torque Driver (DID-05)
Previous | 57 / 113 | Next |
CEDAR - Digital Torque Driver (DID-05)
Brand: Cedar - Japan
Model: DID-05
Features:
Specifications:
Brand: Cedar - Japan
Model: DID-05
Features:
- The torque screwdriver is Ergonomic and light weight, yet ruggedly constructed
- Peak, Real Time and Peak Down measuring mode (selectable)
- 800 data memory
- Programmable High and Low set points with both audible beep and Green/Red LED indicator for uniform torque tightening or GO/NO GO testing
- Programmable screw tightening counter
- Programmable Auto Zero function resets unit to zero for easy operation
- Both CW and CCW operation
- ASCII format output (USB)
- Runs on internal NiCad batteries (for 12 hours) Auto shut-off after 10 min. of non-use
Specifications:
Model Formula | DID-05 | |
Measuring range | 2.0¡«500 mN・m 0.020¡«5 kgf・cm 0.020¡«4.5 lbf・in | |
Accuracy | ±0.5£¥ (499 or less digit±3digit) | |
Display | The 4 figures digital display of LCD | |
The measurement direction | CW-CCW (right and left) | |
Measurement Mode |
PP | peak to peak The highest value of load is hold-displayed. |
TR | track A display also changes with load change. | |
PD |
peak-down Load catches and displays the moment included in descent from a rise. |
|
C |
Real-time Output Load torque value is outputted every about1/ 160 second. |
|
A success or failure judging function |
High and low value are measurement within the limits, and can be set up. Five-set preservation is possible. |
|
The Maximum, minimum, and the average value |
The data number, the maximum, the minimum value, and the average value of memory data are displayed. |
|
Data memory counts | 800 data | |
Data output | ASCII format (baud rate 19200) | |
auto clear time | If the set-up time comes, measurement value will be automatically cleared in 0.0~3.0 seconds (0.5s interval). | |
Setup of count | 1 to 99 count (count in a clockwise direction) | |
Power supply | Ni-Mn chargeable battery 650mAh | |
Charge time / use time | Charge about 3 hour /, continuation about 12-hour use | |
Auto-power-off | After the neglect during 10 minutes | |
Bit / Socket part | φ4 | |
Weight | 180£ç | |
Accessories (one piece each) |
φ4 Bit + #1£¨For one-way×1 For fixation×1) | |
Exclusive charger | ||
Result of calibration document Certification on calibration document / Traceability system figure |
||
Option | When outputting data, the cable (USB miniB) is needed. | |
Outside-dimension (m/m) |
|