- Color & Paint Coating Measurement
- Dimensional Measurement
- Electrical & Electronic Meter
- Electrostatics (ESD) , Ionizer & Resistivity Meter
- Environmental & Weather Aging Testing
- Gas Measurement
- Handling & Cutting Tools
- Hardness Testing
- Laboratory Equipments
- Levelling Angle Gauges
- Magnetic Testing
- Medical Equipment
- Metallography
- Microscope, Magnifier & Visual Inspection
- Paper & Film Packaging Testing
- Pressure & Flow
- Rubber & Glove Testing
- Storage Dry Cabinet
- Temperature, Humidity & Moisture
- Tensile & Tension Torque/Force
- Ultrasonic NDT Testing
- Vise & Holder
- Water Analysis Meter
- Weighing
- Others
RANKCONN - Laser Dust Particle Counter Y09-301AC/DC
Previous | 21 / 807 | Next |
RANKCONN - Laser Dust Particle Counter Y09-301AC/DC
Brand: Rankconn Instruments
Model: Y09-301AC/DC
Technical Specification:
Brand: Rankconn Instruments
Model: Y09-301AC/DC
Technical Specification:
Model | Y09-301AC/DC laser dust particle counter |
Sampling flowrate | 2.83L/min |
Particle size binning | 0.3um,0.5um,1um,3um,5um,10um |
Laser light source | Long-life laser diode with a service life of 10 years |
Imported air pump |
Stable continuous work, energy saving and environmental protection |
Accreditation Standards | GB/T6167-2007, JJF1190-2008 |
Repeatability relative error | <±10% |
Particle concentration indication error | <±15% |
Particle size distribution error | <±30% |
Maximum sampling concentration | 35000 capsules/liter |
Self-cleaning time |
≤ 10min (count 3 times in 10 minutes in a row is zero, 95% confidence) |
Sampling time | User Configurable (5」9999 sec.) |
Sampling delay | User self-set (1」100 seconds) interval 1-1800S |
Number of UCL samples | 1」99 |
Number of UCL sampling points (location) | 2」50 |
Working hours | The continuous test time is about 5 hours |
Data memory capacity |
1」10000 sets of measurement data, which can be queried |
Counting mode | Cumulative |
On-screen display | Color 5-inch touch screen |
Printing method | Built-in thermal printer |
Charging power source | AC 220V/50Hz |
Built-in lithium battery | DC 16.8V 2.2Ah |
Charging adapter | Input 100-240V 50Hz, output 16.8V 2A |
Battery charging time | 6 hours |
Cleanliness level judgment standard |
ISO14644-1, GMP Static Mode, GMP Dynamic Mode |
Temperature and humidity | Optional |
Data communication interface | USB |
Environment |
Operating environment: 0」40°C relative humidity 10」90% |
Storage environment: -30°C」45°C relative humidity 0」90% |
|
Standard accessories |
Isodynamic sampling head, sampling head connected to PVC hose (inner diameter 6mm), self-cleaning filter, tripod, printing paper, portable protective case, charging adapter, U disk, USB data cable, instruction manual, original factory calibration report, certificate |
Dimensions | Length 170 x Width 170 x Height 160 |
Material | SUS304 stainless steel |
Weight (with battery) | 2.4Kg |